Model Number 3100 A
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/12/2020
Event Type  malfunction  
Manufacturer Narrative
Vyaire field service representative (fsr) went onsite and evaluated the ventilator.The tp4 reading on power controller were 3.33 and -4.59 where it should be 3.5 and -3.95.Fsr installed new pot, driver power module and driver assembly.Calibrations were done and passed the test.Fsr run the unit for about 20 minutes on different power settings no errors found and pressure remained constant according to settings.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reported a low delta p on 3100a while connected to a patient.At this time, there is no information regarding patient harm and remedial action taken by the healthcare facility associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, il
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs, ca
Manufacturer Contact
mindy faber
26125 n. riverwoods blvd.
mettawa, il 
MDR Report Key10131369
MDR Text Key194532822
Report Number2021710-2020-12043
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20130501
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/12/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-