ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL
|
Back to Search Results |
|
Model Number 21AGFN-756 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/18/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Further information regarding this event has been requested.If returned, investigation results will be provided in a subsequent submission.
|
|
Event Description
|
On (b)(6) 2020, a sjm regent heart valve w/flex cuff was chosen for procedure.During the procedure, as the user pushed the aortic valve down into the annulus, one of the leaflets broke off the valve.
|
|
Manufacturer Narrative
|
The reported event of one of the leaflets breaking off the valve was confirmed.One leaflet was dislodged from the orifice and returned with the valve.A small chip was noticed in on recessed pivot area.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the leaflet dislodgment could not be conclusively determined.
|
|