• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HP054CN
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/15/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # t92z80.Investigation summary.The hand piece was received with the cable insulation damaged.In addition the housing was not broken nor damaged as reported.However this did not create a functional issue.The device was connected to a generator, evaluated with a test instrument and was found to be functional.The instrument was disassembled to perform an internal electrical test that revealed a hand activation open circuit condition at the cable level, affecting the functionality of the handpiece.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.No conclusion could be reached as to what caused this issue.Possible causes that may have contributed to the cable damage include the cable being ran over, or getting tangled by the cart wheels.A manufacturing record evaluation was performed for the finished device and the manufacturing criteria were met prior to the release of this lot/batch.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that the handle was broken.The inner core of the handle broke.Changed to another one to complete surgery.There was no patient consequence reported.
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC SCALPEL HAND PIECE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key10135669
MDR Text Key194618678
Report Number3005075853-2020-02926
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
PMA/PMN Number
K002906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHP054CN
Device Lot NumberT92Z80021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2020
Date Manufacturer Received05/18/2020
Patient Sequence Number1
Treatment
GENERATOR
-
-