MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GII P/S CONS HOUSING REAM DOME; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GII P/S CONS HOUSING REAM DOME; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, Back to Search Results
Catalog Number 71440145
Device Problem Product Quality Problem (1506)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/14/2020
Event Type  malfunction  
Event Description
It was reported that during inspection was found that the gii p/s cons housing ream dome is dull.There was no case involved.
 
Manufacturer Narrative
Find additional information received for this case in sections: d4 (lot number and expiration date) and h4.
 
Manufacturer Narrative
Results of investigation: the device, intended for use in treatment, was returned for evaluation.A visual inspection was conducted and confirmed the housing reamer dome is dull.The device was manufactured in 2012.This device shows signs of extreme wear/usage.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.This device is reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII P/S CONS HOUSING REAM DOME
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10138578
MDR Text Key194650417
Report Number1020279-2020-02298
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71440145
Device Lot Number12MM10858
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2020
Date Manufacturer Received08/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-