• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFC STEM TRIAL EXTRACT; KNEE INSTRUMENT : EXTRACTION INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFC STEM TRIAL EXTRACT; KNEE INSTRUMENT : EXTRACTION INSTRUMENT Back to Search Results
Model Number 86-5226
Device Problem Break (1069)
Patient Problems Not Applicable (3189); No Code Available (3191)
Event Date 06/02/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Revision total knee; the stem trial extractor was broken whilst trying to remove the fluted rod trial.The tip broke & the stem trial had to be removed with other general surgical instruments.No pieces were left in the patient.No ae to patient.7 minute delay to the procedure.
 
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned device confirmed the reported event.The noted damage is consistent with improper user technique through excessive leveraging/torquing of the device and the investigation did not establish a need for corrective action.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Event Description
Additional information received stated that the piece that broke off the stem extractor remained in the fluted stem trial.The stem trial connection to the drill wasn¿t as snug as a new instrument.
 
Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: corrected h3.Added d10.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC STEM TRIAL EXTRACT
Type of Device
KNEE INSTRUMENT : EXTRACTION INSTRUMENT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10145358
MDR Text Key194958105
Report Number1818910-2020-13678
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10603295216766
UDI-Public10603295216766
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 06/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number86-5226
Device Catalogue Number865226
Device Lot NumberTX6515
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2020
Date Manufacturer Received07/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-