• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX 15CMX10M CTN 1; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. OPSITE FLEXIFIX 15CMX10M CTN 1; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Catalog Number 66000375
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/14/2020
Event Type  malfunction  
Event Description
It was reported that the dressings are not sticky enough for being used.No harm or injury reported.
 
Manufacturer Narrative
H10.H3, h6: we have now concluded our investigation for the complaint received.A review of the associated batch manufacturing records did not identify any issue at the point of manufacture which may have caused or contributed to the issue highlighted by the complainant.In addition, it can be confirmed that all finished product specification testing was satisfied at the point of release.A complaint history review was carried out using the lot and part numbers provided, there have been no further complaints reported with this failure mode in the past three years.The device is intended for use in treatment.As no samples were returned a thorough product evaluation could not be carried out.However, an image provided demonstrated the reported issue.This confirmed a relationship between the event and the device.Low adhesion may be caused by adhesion levels of the raw material used to make the film of this dressing.There are various controls in place during the manufacturing process of both the raw material and the dressing to identify when the adhesion is not in acceptable range, as per the product specification.There are also regular tests carried out to monitor the finished product.The root cause was traced to manufacturing.The current occurrence rate is within our acceptable range, therefore, no further actions by smith + nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith + nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPSITE FLEXIFIX 15CMX10M CTN 1
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10145935
MDR Text Key194877319
Report Number8043484-2020-00686
Device Sequence Number1
Product Code KGX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 08/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number66000375
Device Lot Number1827
Was Device Available for Evaluation? No
Date Manufacturer Received08/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-