• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Degraded (1153)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/19/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported the threads on the trial gauge inserter are damaged, it will not screw on.No patient impact or delay in surgery.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event Description
No additional information on reported event.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g4; h2; h3; h6 complaint sample was evaluated and the reported event was confirmed.Inspection of the returned device revealed shaving within the threads of the inserter.Assembled photos 1 and 2 show the ends of the shaving and the shaving wrapped around the tip.During the evaluation, the shaving was able to be unscrewed from the inserter tip.Visual inspection could not confirm if the shaving is part of the inserter or a mating device, although the shaving is likely from a different device.The threads of the inserter are deformed.The shaft is scratched and discolored consistent with a multiple use instrument.Impact marks were observed on the strike plate.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10147839
MDR Text Key194944201
Report Number0001825034-2020-02279
Device Sequence Number1
Product Code PBI
UDI-Device Identifier00880304532694
UDI-Public(01)00880304532694
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 07/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number554480
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2020
Was the Report Sent to FDA? No
Date Manufacturer Received07/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberNI
Patient Sequence Number1
-
-