Model Number N/A |
Device Problem
Degraded (1153)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/19/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported the threads on the trial gauge inserter are damaged, it will not screw on.No patient impact or delay in surgery.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No additional information on reported event.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g4; h2; h3; h6 complaint sample was evaluated and the reported event was confirmed.Inspection of the returned device revealed shaving within the threads of the inserter.Assembled photos 1 and 2 show the ends of the shaving and the shaving wrapped around the tip.During the evaluation, the shaving was able to be unscrewed from the inserter tip.Visual inspection could not confirm if the shaving is part of the inserter or a mating device, although the shaving is likely from a different device.The threads of the inserter are deformed.The shaft is scratched and discolored consistent with a multiple use instrument.Impact marks were observed on the strike plate.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Search Alerts/Recalls
|