• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARTICULAR SURFACE SIZE 2 9 MM HEIGHT; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARTICULAR SURFACE SIZE 2 9 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problems Positioning Failure (1158); Difficult to Insert (1316)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/20/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source - (b)(6).Customer has indicated that the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during the operation of zuk single condyle replacement, the surgeon performed the osteotomy according to the technical requirements.However, when opening the articular surface, the implant could not be implanted normally.After the attempt, the implant could not be implanted, so the surgeon had to open a box of new implants of the same type again to complete the operation.There is no additional information available at this time.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, d10, g4, g7, h1, h2, h3, h6, and h10.Visual examination of the returned device found damage of the proximal and distal surfaces and damage to the locking feature.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE SIZE 2 9 MM HEIGHT
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10148478
MDR Text Key194967494
Report Number0001822565-2020-02071
Device Sequence Number1
Product Code HSX
UDI-Device Identifier00889024193918
UDI-Public(01)00889024193918
Combination Product (y/n)N
PMA/PMN Number
K033363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 07/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00584202209
Device Lot Number63975407
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2020
Was the Report Sent to FDA? No
Date Manufacturer Received07/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-