BAXTER HEALTHCARE CORPORATION CARTRIDGE; SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE
|
Back to Search Results |
|
Catalog Number 103401 |
Device Problems
Leak/Splash (1354); Improper or Incorrect Procedure or Method (2017)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 05/21/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Initial reporter: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during dialysis treatment with a gambro cartridge set, an external blood leak was observed from the arterial injection site.The volume of the blood loss was not reported.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.A sample analysis was performed and the reported problem was confirmed.Pictures from the attachment showed a leak from crack at the female connector on the service line, indicating the application of excessive force which is likely to cause the reported break.The cause is user error related.Should additional relevant information become available, a supplemental report will be submitted.
|
|