Catalog Number 71440472
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/19/2020
Event Type  malfunction  
Event Description
It was reported that after procedure, the screw came off during cleaning.No pieces fell inside the patient.
 
Manufacturer Narrative
The device, intended use in treatment, was not returned for evaluation, the reported event could not be confirmed.Therefore, product analysis could not be performed at this time.A review of the manufacturing record did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Without the actual product involved, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII TIB TRY IMPL IMPACTOR
Type of Device
TRAY, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10155984
MDR Text Key195233180
Report Number1020279-2020-02413
Device Sequence Number1
Product Code FSM
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71440472
Device Lot Number15DSY0007
Was Device Available for Evaluation? No
Date Manufacturer Received09/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-