The reported event was confirmed, however the cause was unknown.Visual inspection noted that one irrigation syringe was received without the original packaging.Visual evaluation noted that 3 white specs (measuring 0.20, 0.25, 0.20 sq.Mm) were found in the bulb outside the syringe body.Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure mode could be due to defective/ contaminated components from suppliers.The product was not used for diagnosis or treatment.The product had failed to meet the specifications, and was influenced by the reported failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The labelling review was not performed due to labeling could not have prevented the reported failure.
|