• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ACTIS CURVED INSERTER SHAFT; INSERTION DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ACTIS CURVED INSERTER SHAFT; INSERTION DEVICES Back to Search Results
Model Number 2010-07-120
Device Problems Device-Device Incompatibility (2919); Material Twisted/Bent (2981); Mechanical Jam (2983); Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/03/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #
=
> (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the universal stem handle and curved actis stem inserter getting stuck when put together and taken apart.No surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIS CURVED INSERTER SHAFT
Type of Device
INSERTION DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10162378
MDR Text Key195391944
Report Number1818910-2020-13954
Device Sequence Number1
Product Code KWL
UDI-Device Identifier10603295394006
UDI-Public10603295394006
Combination Product (y/n)N
PMA/PMN Number
K150862
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2010-07-120
Device Catalogue Number201007120
Device Lot NumberSO2033822
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2020
Date Manufacturer Received06/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age60 YR
-
-