Patient underload intensive care unit, blood backflow in the substitute at the start of an exchange with comtec.The pump insert (the white one) is set up upside down.Received 2 pictures, on the pictures is clearly visible that the connection of the pump is exchanged.Identified root cause: assembly failure.Result of risk assessment: potential hazard / hazardous situation: tube glued into wrong port of the pump adapter patient underload.Severity of the harm: catastrophic.Occurrence of the harm: improbable.
|