STRYKER ORTHOPAEDICS-MAHWAH NO 3. TRIATHLON TS PLUS TIBIAL INSERT X3 POLY 25MM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
Model Number 5537-G-325 |
Device Problem
Device Difficult to Setup or Prepare (1487)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
|
Event Description
|
It was reported that during a revision of the patient's right knee, the locking wire of the insert came out while positioning the insert (caught on the antero-lateral clip of the baseplate).As no replacement was available in the operating room, the surgeon decided to implant the now wire-less insert and inserted the stabilizer pin further into the baseplate.Surgery was completed successfully with a delay of approximately 5 minutes.Rep confirmed that no further information will be released by the hospital or surgeon.
|
|
Search Alerts/Recalls
|
|
|