Model Number 1565068
Device Problems Degraded (1153); Flaked (1246); Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Date 06/10/2020
Event Type  malfunction  
Manufacturer Narrative
During inspection, a trumpf medical service technician found peeling paint on the yoke arm.The yoke arm was replaced and the unit is functioning as designed.The investigation identified steps in the preparation for coating process which can lead to lack of paint adhesion resulting in paint chipping.Additionally, collisions of the light heads and use of unapproved cleaners can further contribute to paint chipping from the light system.The damage to the painted surfaces usually does not develop suddenly but develops over time.The instructions for use state that the devices must be checked for proper condition prior to each use.Damaged devices must not be used.If damage to the surface coating is recognized and removed, there is no danger to the patient.The surface coating process has been updated to include additional testing and verification.Based on this information, no further action is required.
 
Event Description
The customer reported peeling paint on the yoke arm.No injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILED SINGLE TUSM
Type of Device
SURGICAL LIGHT
Manufacturer (Section D)
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
carl-zeiss strasse 7-9
saalfeld, 07318
GM  07318
Manufacturer Contact
steffen ulbrich
carl-zeiss strasse 7-9
saalfeld, 07318
GM   07318
MDR Report Key10167221
MDR Text Key197354189
Report Number9681407-2020-00037
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102758
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1565068
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/10/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/16/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse