• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HP054
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/27/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation summary the handpiece was received with the nose cone cracked.It was connected to a generator, evaluated with a test instrument and was found to be non-functional.The instrument was disassembled to inspect the internal components and the moisture indicator was positive.Due to the cracked nose cone, moisture entered the hand piece mid housing.Analysis was unable to determine the exact root cause that lead to crack in the hand piece nosecone.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.It is possible that the ingress of moisture affected handpiece functionality.
 
Event Description
It was reported that the device did not work despite 33 uses remained.No patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC SCALPEL HAND PIECE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 00969
*   00969
6107428552
MDR Report Key10167694
MDR Text Key195567230
Report Number3005075853-2020-03076
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036002147
UDI-Public10705036002147
Combination Product (y/n)N
PMA/PMN Number
K002906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other
Type of Report Initial
Report Date 05/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHP054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2020
Date Manufacturer Received05/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
GENERATOR.
-
-