The reporter's meters were returned for investigation.The test strips were not returned.The meters were tested with retention test strips lot 478646 exp 30-jun-2021: testing results for uu14390008: control ranges: sample 1: 2.8-4.0 mmol/l.Sample 2: 9.5-11.6 mmol/l.Sample 3: 19.5-23.8 mmol/l.Results: sample 1: 4.1, 4.1, 3.9 mmol/l.Sample 2: 10.4, 10.3, 10.4 mmol/l.Sample 3: 21.4, 21.8, 21.4 mmol/l.Testing results for uu14379836: control ranges: sample 1: 2.8-4.0 mmol/l.Sample 2: 9.5-11.6 mmol/l.Sample 3: 19.5-23.8 mmol/l.Results: sample 1: 4.0, 4.0, 3.9 mmol/l.Sample 2: 10.4, 10.2, 10.4 mmol/l.Sample 3: 21.7, 21.8, 22.2 mmol/l.Testing results for uu14189223: control ranges: sample 1: 2.8-4.0 mmol/l.Sample 2: 9.5-11.6 mmol/l.Sample 3: 19.5-23.8 mmol/l.Results: sample 1: 3.8, 3.8, 3.9 mmol/l.Sample 2: 10.3, 10.3, 10.3 mmol/l.Sample 3: 21.5, 21.5, 21.6 mmol/l.All returned results are within acceptable range.No information was provided in the complaint case that would point to a cause for the result discrepancy.
|