Model Number 2274-52-000
Device Problems Break (1069); Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/15/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Posterior corial pinnacle during surgery rep present surgeon reported that drill bit broke when removing from the wound, whilst drilling for a cancellous screw with a gription cup already in situ.Rep then identified a number of other drill bits missing from the consignment tray, which have also been listed for further consignment replacements.Simple replacement of consignment inventory required - no delay to surgery.
 
Event Description
The middle part of the shaft broke, coiled shaft.
 
Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET FLEX DR SFT QCK CPLE
Type of Device
HIP INSTRUMENTS : INSERTION DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10176594
MDR Text Key195908942
Report Number1818910-2020-14199
Device Sequence Number1
Product Code HTW
UDI-Device Identifier10603295109273
UDI-Public10603295109273
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 06/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2274-52-000
Device Catalogue Number227452000
Was Device Available for Evaluation? No
Date Manufacturer Received07/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-