This follow-up report is being submitted to relay additional information.The following sections were updated: b4, d10, g4, g7, h2, h3, h6, h10.The complaint sample was evaluated and the reported even was confirmed.The returned product packaging showed foreign matter within the sterile barrier.The debris was evaluated and found to be non-conforming.The device history records were reviewed and no discrepancies were identified.The device was likely non-conforming when it left zimmer biomet control.The root cause of the reported event can be attributed to the operator not following instructions during manufacturing.No further actions have been initiated to further address the reported issue.This complaint was determined not to be a new confirmed quality or manufacturing issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|