Device evaluation: b1, b2, d10, g4, g7, h1, h2, h3, h6 and h10 device evaluation: the vyaire failure analysis laboratory received the suspect component and performed a failure investigation.The issue reported by the customer could not be duplicated.The assembly was installed into a known good unit and no fluctuations were found during the operation.The system was left to cycle overnight and the readings did not change.The panel meter was sent to the manufacturing line for testing and passed with a reading of 50.0 which is what is expected.Additional information: the patient was removed from the ventilator.The customer confirmed that there was no patient harm associated with the reported event.
|