The device was returned for analysis.The tip, device shaft and sensor port were examined for damage or any irregularities.The wire shaft showed a kink at the proximal end of the wire where the wire connects to the occ handle.There was a kink located at 38cm from the tip.The tip showed bend damage and stretched coils.Since no occ handle was received for analysis, a test occ handle was used for functionality of the wire.The wire was connected to the test occ handle.The occ handle was connected to the ffr link to verify the signal strength.The signal was present and showed green lights as designed.The occ handle was then connected to the bench top testing equipment and the wire was inserted into the pressure chamber.The pressure was increased to verify the sensor was indeed reacting to the pressure increases and decreases.The pressure sensor functioned as designed.The coefficient values could not be verified due to the occ handle not being returned.The occ handle was again connected to the ffr link.The device was then connected to the polaris (ilab) test equipment via bluetooth signal.The wire communicated to the polaris system and zeroed as designed.With the wire inserted into the test pressure chamber, the wire transferred a pressure waveform to the polaris which indicates a functioning wire.The wire was removed from the occ handle.The sensor port showed no residue of body fluids.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|