Model Number 37603 |
Device Problems
Shipping Damage or Problem (1570); Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 37603, serial#: (b)(4), product type: implantable neurostimulator.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the battery voltage was lower than usual before implantation.The battery voltage was 2.98v.It was unknown if it was a defect but was not used.Unknown if any environmental/external/patient factors led or contributed to the issue.No diagnostics/troubleshooting was done.Another product was used.No patient symptoms were reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received stating that the product was retrieved and would be sent to the analysis center.It was requested to discard the defective product.
|
|
Event Description
|
Additional information was received from the manufacturer representative (rep) stating that the cause of the battery voltage issue was not determined.It was also stated that the battery voltage issue occurred just before taking the product out of the box.The issue was resolved after another implantable neurostimulator (ins) was used.The defect product was retrieved but the customer requested to discard it.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Analysis of the implantable neurostimulator (ins) (serial #(b)(6)) revealed that no anomalies were identified and the device passed all testing.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|