This supplemental report is being submitted to provide additional information.Olympus medical systems corp.(omsc) could not investigate the subject device, because the subject device was not returned to omsc.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.A potential cause of the event is as follows.The customer had stored the endoscope by placing it in the reprocessing basin of the olympus endoscope reprocessing machine, oer-s.The reprocessing basin has high humidity and microorganisms are easy to grow in the environment.The endoscope was likely to be contaminated when being stored in the basin.In addition, reportedly, the user didn¿t check the concentration of the disinfectant solution or change disinfectant solution in accordance with the instruction manual.
|