• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER ADVANCED DSP TIP SCISSORS; SCISSORS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER ADVANCED DSP TIP SCISSORS; SCISSORS, OPHTHALMIC Back to Search Results
Catalog Number 725.52P
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/12/2020
Event Type  malfunction  
Manufacturer Narrative
No sample has been received by manufacturing for evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A nurse reported that an ophthalmic scissors tip broke while in contact with the patient during vitrectomy surgery.There was no harm to the patient.Additional information received further clarified that the tip did break while inside of the patient's eye, but the broken tip was successfully removed from the eye.
 
Manufacturer Narrative
The received sample was found in the inner and outer blister without the cover foil.A review of the device history record traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria.The investigated instrument had one scissor blade broken off.As it was reported, by the customer.The instrument was investigated.The fracture surface did not show signs of corrosion.Further signs indicating, the cause of the fracture were not found, during the investigation.A 100% functional test is performed, during production.Which ensures that the instrument fulfills the specification, before it will be delivered.Therefore, it can only be assumed that issue was caused afterwards during handling, but this can not be verified.The root cause was unable to be verified as no signs were found, indicating the cause of the fracture.A manufacturing or design related root cause for the damage of the complained device has not been identified.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken, as appropriate.At a minimum, this will include completing reviews of complaint class report levels on a monthly basis.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER ADVANCED DSP TIP SCISSORS
Type of Device
SCISSORS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
MDR Report Key10200328
MDR Text Key197103979
Report Number3003398873-2020-00048
Device Sequence Number1
Product Code HNF
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Catalogue Number725.52P
Device Lot NumberF187002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Date Manufacturer Received03/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-