• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FEMORAL HEAD/NECK IMPACTOR; PROSTHESIS,HIP,SEMI-CNSTRND,UNCEMENTED,METAL/PLYMR,NON-POROUS,CALICUM-PHSPHT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FEMORAL HEAD/NECK IMPACTOR; PROSTHESIS,HIP,SEMI-CNSTRND,UNCEMENTED,METAL/PLYMR,NON-POROUS,CALICUM-PHSPHT Back to Search Results
Catalog Number 71360093
Device Problem Break (1069)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/05/2020
Event Type  malfunction  
Event Description
It was reported that during efip inspection the plastic tip was found broke off.No patient involvement.
 
Manufacturer Narrative
The device, intended for use in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is broken and has signs of wear and tear from use.The device was manufactured in 2015.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD/NECK IMPACTOR
Type of Device
PROSTHESIS,HIP,SEMI-CNSTRND,UNCEMENTED,METAL/PLYMR,NON-POROUS,CALICUM-PHSPHT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10210233
MDR Text Key196910888
Report Number1020279-2020-02835
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71360093
Device Lot Number15CM18843
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2020
Date Manufacturer Received12/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-