Upon root cause analysis by our service technician, it was found that the daily calibration had failed and the customer cleared the message and scanned anyways.It was seen that the workstation sent the scanner the notification but images were not in the browser or in the data folder on the workstation.The customer resent the study from the ui and the images were loaded.A tick calibration was completed and the raw data was checked.A daily cal and qa was completed as well.Device is operational and was returned in to service.It was found that the patient received the following dose amount:: ctdivol (mgy) : 0.623, dlp (mgy.Cm) : 31.16.
|