RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Catalog Number 261221
Device Problem Activation Problem (4042)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2020
Event Type  malfunction  
Manufacturer Narrative
Device identifier # (b)(4).Perforator was not returned for evaluation; therefore, an evaluation of the device could not be performed.Lot number information has been provided; therefore, manufacturing records were reviewed and found no anomalies.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.
 
Event Description
A facility reported that the surgeon complained that the codman disposable perforator did not stop and plunged through the dura while doing a burrhole for drainage of a chronic subdural.The surgeon had to pull out the perforator.There was no direct injury to the brain.No surgical delay was noted.
 
Event Description
N/a.
 
Manufacturer Narrative
Device identifier # (b)(4).The perforator was returned for evaluation: device history record (dhr): there is no indication that the production process may have contributed to this complaint.Failure analysis: visually inspected utilizing unaided eye: organic matter and a wrong eto label was found.The complaint could not be verified through failure analysis.Tested with in the specifications (ifu testing, functional testing).The root cause is undetermined and was unable to be confirmed in the complaint evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA 02767
MDR Report Key10221598
MDR Text Key200774749
Report Number1226348-2020-00354
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 06/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2020
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number261221
Device Lot NumberJ57D54
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2020
Date Manufacturer Received08/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-