Model Number PFSR110137
Device Problems Break (1069); Material Separation (1562)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2020
Event Type  malfunction  
Event Description
It was reported that during the testing of the machine in a lab/demo, it was found that the snaplock nut of the handpiece was broken.No patient was involved.No other complications were reported.
 
Manufacturer Narrative
H3, h6: the navio handpiece, part pfsr110137 intended for use in treatment was not made available to the designated complaint unit for evaluation thus, a visual and functional evaluation could not be performed.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, a contributing factor may be damaged or broken component.No containment or corrective actions are recommended at this time.If the product associated with this event is returned or provided at a future date, this evaluation will be reopened for investigation.
 
Manufacturer Narrative
H10: internal complaint reference (b)(4).The navio handpiece, part number pfsr110137, serial (b)(6) and intended to be used for treatment, was returned for evaluation.A relationship between the reported event and the device was established.The reported problem could not be visually confirmed.A functional evaluation was performed.The reported problem was confirmed.Upon turning the snaplock nut, it was found to easily separate from the snaplock thread receiver.A review of manufacturing records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most probable cause of the reported problem is mechanical failure with the snaplock component.The reported failure was assessed in a capa investigation.As a response to the capa, the snaplock nut underwent a revision to include a pin which is intended to fixate the nut in place with the snaplock thread receiver.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
suite 100
austin, TX 78735
5123913905
MDR Report Key10228604
MDR Text Key197580209
Report Number3010266064-2020-01639
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-