Catalog Number A1114 |
Device Problems
Mechanical Problem (1384); Material Separation (1562); Naturally Worn (2988)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not yet received by the manufacturer for evaluation.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
|
|
Event Description
|
A customer reported that the triad rocker arm attachment of the a1114 mayfield infinity skull clamp started to separate when pressure was added during the placement process leading to the locking knob being unable to unlock.It was also reported that the starbursts were worn down.There was no known patient injury or delay in surgery.
|
|
Manufacturer Narrative
|
(b)(4).The device was returned for evaluation.The device history record (dhr) showed no abnormalities related to the reported failure.The device passed all required dimension check and inspection points with no associated mrr¿s, variances or rework.The evaluation showed that the triad rocker arm was functioning properly when pressure was added.The observed condition is likely caused by wear and tear or improperly handling of the device.The definite root cause cannot be reliably determined.The reported compliant was not confirmed from the evaluation.No issues observed from the evaluation.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|
Event Description
|
N/a.
|
|