• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS MODUTEC; TUBING, PRESSURE AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS MODUTEC; TUBING, PRESSURE AND ACCESSORIES Back to Search Results
Device Problem Crack (1135)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/19/2020
Event Type  malfunction  
Manufacturer Narrative
Issue is being investigated by manufacturing site.Device not returned to manufacturer.
 
Event Description
On 1st july, 2020 getinge became aware of an issue with one of devices- modutec.As it was stated, the seal was found cracked.There was no injury reported however we decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.
 
Manufacturer Narrative
Getinge became aware of an issue with one of devices- modutec.As it was stated, the seal was found cracked.There was no injury reported, however, we decided to report the issue in abundance of caution as any particles falling off into sterile field or during procedure may cause contamination.It was established that when the event occurred, the device did not meet its specification as the seal should not crack and it contributed to incident.There is no information if when the event occurred the device was or was not being used for patient treatment.The tearing and cracks observed on the rubber¿s bellows are probably due to the age.According to their aspect they are probably very old (more than 10 years).They should have been changed earlier because this type of damages appears over time and can easily be detected during preventive maintenance.A preventive maintenance carried out every year would avoid the complete tearing.We believe that if the manufacturer recommendation would have been followed the incident would have been avoided.Given the circumstances we shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Event Description
Manufacturer reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODUTEC
Type of Device
TUBING, PRESSURE AND ACCESSORIES
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
MDR Report Key10243935
MDR Text Key198190908
Report Number9710055-2020-00209
Device Sequence Number1
Product Code BYX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-