SMITH & NEPHEW, INC. UNKNOWN TANDEM BIPOLAR FEM HEAD; PROSTHESIS,HIP,HEMI-,TRUNNION-BEARING,FEMORAL,METAL/POLY
|
Back to Search Results |
|
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/10/2020 |
Event Type
Injury
|
Event Description
|
It was reported that during reduction after implanting, head and stem were dislocated.Operative duration was extended less than or equal to 30 minutes.
|
|
Manufacturer Narrative
|
The device, used in treatment, was not returned for evaluation, therefore the failure could not be confirmed.The clinical/medical team concluded, this complaint from japan reports that during the primary surgery after the implanting the head and stem were found to be dislocated.The ¿operative duration was extended¿ ¿less than or equal to 30 minutes.¿ it was communicated that no further medical documentation would be forthcoming.Smith and nephew has not received adequate materials( the operative reports) to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to traumatic injury, patient anatomy or abnormal loading of limb.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|