Results of investigation: the device, used in treatment, was returned for evaluation.A visual inspection confirms the entry point of the pin drivers were the connection of the pin goes is severely damaged.This device shows significant signs of wear/usage.This device was manufactured in 2013.A functional evaluation confirmed the stated failure.The driver would not hold a.123 pin gage as intended.A medical investigation will be performed.Proceed based on information provided / available for the investigation; if no relevant clinical information is provided, recommend closure.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear / tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.
|