• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXF TRL BRG W/SLOTS MED 4MM; KNEE TIBIA TRIAL PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF TRL BRG W/SLOTS MED 4MM; KNEE TIBIA TRIAL PROSTHESIS Back to Search Results
Model Number N/A
Device Problems Break (1069); Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/17/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it have been discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the oxford uni size medium 4mm thickness trial bearing broke as surgeon tries to extract it from patient.No delay of the surgery reported.
 
Manufacturer Narrative
(b)(4).As the product has not been received, the investigation was limited to the information provided.A review of the manufacturing history records was not performed as lot number is not available.A review of the complaint database over the last 3 years has found no similar complaints reported with these items.Without the opportunity to examine the complaint product, the root cause cannot be determined due to insufficient information.Risk assessment: risk management report documents the estimated residual risk associated with the reported event.The reported event states trial bearing fracture.The details of the reported event do not allege that there was any patient harm or delay to surgery.Therefore, the severity is considered s-1 (negligible) as per definitions within the severity table in the rmr, which is in line with the risk file.The outcome of this complaint is considered to be within the severity of the rmr.No corrective or preventive actions are deemed necessary at this time.If any additional information becomes available, then the complaint will be reopened and investigated thoroughly.
 
Event Description
It was reported that the oxford uni size medium 4mm thickness trial bearing broke as surgeon tries to extract it from patient.No delay of the surgery reported.No harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF TRL BRG W/SLOTS MED 4MM
Type of Device
KNEE TIBIA TRIAL PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
MDR Report Key10263051
MDR Text Key199908551
Report Number3002806535-2020-00318
Device Sequence Number1
Product Code NRA
Combination Product (y/n)N
PMA/PMN Number
P010014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number32-422704
Device Lot NumberN/A
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-