MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS SYSTEM; PROTHROMBIN TIME MONITOR Back to Search Results
Model Number XS PST
Device Problem Circuit Failure (1089)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/10/2020
Event Type  malfunction  
Manufacturer Narrative
Occupation is lay user/patient.The meter was requested for investigation.
 
Event Description
The initial reporter complained of an issue with a coaguchek xs meter.The customer stated there was a brown spot in the upper left corner of the display screen; the customer stated it looked like a spot on the meter screen was burned.The customer did not see the brown spot on the screen when the meter was turned on; the brown spot was only visible when the meter was turned off.The customer took the batteries out of the meter "last week" and noticed a black, rubbery material inside the battery compartment.
 
Manufacturer Narrative
The customer's meter was received for investigation.The meter¿s circuit board was tested for damage or contamination and it was found to be free of contamination and damage.The display functionality was checked and the display showed no defects or malfunctions.Medwatch fields d10 and h3 have been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS SYSTEM
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key10263809
MDR Text Key199909801
Report Number1823260-2020-01669
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 08/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/13/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberXS PST
Device Catalogue Number03666794001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2020
Date Manufacturer Received06/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-