BOSTON SCIENTIFIC CORPORATION AXIOS STENT AND DELIVERY SYSTEM; PANCREATIC STENT, COVERED, METALLIC, REMOVABLE
|
Back to Search Results |
|
Model Number M00553560 |
Device Problem
Unsealed Device Packaging (1444)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/01/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event was approximated to (b)(6) 2020 as no specific event date was reported.The complainant was unable to provide the suspect device lot number.Therefore, the manufacture and expiration dates are unknown.(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation on june 22, 2020 that a hot axios stent was to be used to treat a walled off necrosis during an endoscopic ultrasound (eus) with stent placement procedure performed on an unknown date.According to the complainant, during preparation, it was noted that the inner sterile pouch was not sealed properly.Another hot axios stent was used to complete the procedure.There were no patient complications reported as a result of this event.
|
|