|
Catalog Number 42555 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 06/26/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the user was unable to connect the nipple & nut to an oxygen tube as the connection was loose.Therefore, a new unit was used instead.No patient harm reported.
|
|
Manufacturer Narrative
|
Qn# (b)(4).The sample was returned for evaluation.A visual inspection was performed and it was observed that the assembly of the nut and nipple is loose as described in the complaint.Damage was found on the internal tabs of the nut.No other issues were found.The device history record of batch number 74h1900109 was reviewed and no issues or discrepancies were found which could potentially be related to this complaint.No non conformance reports were originated for the lot in question that can be associated to the complaint reported.The dhr shows that the product was assembled and inspected according to specifications.Based on the investigation performed, the reported complaint is confirmed.Although the complaint is confirmed , there is not e nough evidence to assure that this issue was originated during the manufacturing assembly or molding process.The damage on the internal tabs of nut were most likely caused by the end user during the connection of the nut into the flowmeter.The root cause for the condition reported could not be identified.
|
|
Event Description
|
It was reported that the user was unable to connect the nipple & nut to an oxygen tube as the connection was loose.Therefore, a new unit was used instead.No patient harm reported.
|
|
Search Alerts/Recalls
|
|
|