• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 21AGFN-756
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/11/2020
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2020, a 21 mm regent heart valve was selected for implant.While placing the valve the physician adjusted the direction of the leaflet and it fell off.The valve was replaced with a 19 mm regent heart valve and due to the need of a new valve there was a clinically significant delay in procedure.The patient was report to be in stable condition.
 
Manufacturer Narrative
The reported event of a dislodged leaflet was confirmed.Both leaflets were dislodged from the orifice, with one leaflet returned with the valve.That leaflet was fractured, and one pivot guard was damaged.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the leaflet dislodgment and fractured pivot guard could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key10270249
MDR Text Key198700335
Report Number3003681312-2020-00008
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21AGFN-756
Device Catalogue Number21AGFN-756
Device Lot Number7392486
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2020
Was the Report Sent to FDA? No
Date Manufacturer Received07/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-