Model Number N/A |
Device Problem
Device Damaged by Another Device (2915)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/17/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported the shell impactor screw threads broke while trying to place the cup during an initial tha.A backup impactor was used with no further issues.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the g7 straight monoblock shell inserter confirmed the threaded tip to be fractured.A portion of the face of the tip remains intact.Scratching and scuffing consistent with a multiple use device was observed on the shaft and impact marks were observed on the strike plate.(b)(4) an xrf analysis confirmed the inserter material to be consistent with 455 stainless steel alloy.Conclusions: the common failure mode for the threaded inserter tips presented in this summary is bending overload of the threaded tip, some of which may have also had minimal threads engaged at the time of fracture.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined, however sem analysis suggests bending overload may have caused or contributed to the fracture of the threads.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional information on the reported event.
|
|