INTEGRA LIFESCIENCES CORPORATION DURAGEN 3X3 1 PACK CE; N/A Back to Search Results
Catalog Number ID3301I
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Adhesion(s) (1695); Laceration(s) (1946)
Event Date 11/15/2019
Event Type  Injury  
Manufacturer Narrative
The device is not expected to be returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.Linked to mfg report no.: 1121308-2019-00048.
 
Event Description
This is 1 of 2 reports.A distributor reported on behalf of the customer that both id4501i duragen 4x5 ("primary product") and id3301i duragen 3x3 ("combination product") were used on (b)(6) 2019 for an external decompression for obstructive hydrocephalus due to hemorrhagic cerebral infarction and brain hernia.On (b)(6) 2019, a craniotomy and scalp detachment procedure were performed.They found severe adhesions between the scalp and regenerated dura where the duragens were placed.The male patient had incurred a dural laceration due to the surgery to remove the adhesions.The cranioplasty was delayed for over 30 minutes.New duragen was used on the patient for the re-dura formation.The patient was being closely monitored.
 
Manufacturer Narrative
The device was not returned for evaluation.The dhr review was not possible because the lot numbers for the catalogs reported were not provided by the customer.The complaint is considered unconfirmed.The root cause for this event is unknown.Between 05nov2019 and 30jun2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from plainsboro, nj to princeton, nj.Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa (b)(4) have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period 05nov2019 through 30jun2020.Integra lifesciences contacted (b)(6) mdr team, office of product evaluation and quality on july 8-9, 2020 to report these issues regarding mdr reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURAGEN 3X3 1 PACK CE
Type of Device
N/A
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION
105 morgan lane
105 morgan lane
plainsboro NJ 08536
MDR Report Key10276238
MDR Text Key204049193
Report Number1121308-2019-00047
Device Sequence Number1
Product Code GXQ
Combination Product (y/n)N
PMA/PMN Number
K982180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Type of Report Initial,Followup
Report Date 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberID3301I
Was Device Available for Evaluation? No
Date Manufacturer Received12/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-