INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problem Decrease in Pressure (1490)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/22/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.Linked to mfg report no.: 3004608878-2020-00093.
 
Event Description
This is 1 of 2 reports.A senior neurosurgery sales specialist reported that on (b)(6) 2020 the a1059 mayfield modified skull clamp was used in a surgery and the surgeon positioned the clamp with 60lbs of pressure on the torque knob and then shortly thereafter noticed that the pressure had dropped to 40lbs.There was no pin slip or laceration to the patient.The surgeon increased pressure back to 60lbs and proceeded with the procedure.Additional information has been received on 24jan2020 that there was no revision/medical intervention required since they just screwed the device down to 60lbs and started the case with no delay.Additional information has been requested.
 
Manufacturer Narrative
Additional information: d10, g4, g7, h2, h3, h6, h10 device identifier # (b)(4) the device was not returned for evaluation therefore the failure analysis to identify root cause to the end users experienced could not be determined.No dhr review was possible as no lot or serial number was provided.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.Additional information received on 29jan2020 stated that integra disposable skull pins were used on both clamps during procedure.Between 05nov2019 and 30jun2020, approximately 2,200 mdrs submitted electronically by integra lifesciences via trackwise, integra's complaint handling system, were not received by cdrh due to a computer system issue.Within this time period, an error with integra's middleware, which facilitates communications between trackwise and the fda system, caused the complaint records to close and indicate we had received an acknowledgement 3 from the fda when we had not.Integra interpreted the acknowledgement as a successful submission; however, subsequent investigation revealed the acknowledgement 3 received was from our middleware and not from the fda (these acknowledgements have been retained as part of the documentation of the mdr).The malfunction was related to the relocation of the trackwise application to a new data center during the transition of integra's corporate headquarters from plainsboro, nj to princeton, nj.Previously, integra had been successfully receiving acknowledgements 1, 2, and 3 from the fda, and our records reflect these acknowledgements, including the date and time stamps.Capa pr 229048 and nc 20-011 have been opened by integra to further investigate the nonconformance and develop a corrective action plan.The middleware error has been corrected, and integra has filed mdrs since the correction and verified that the appropriate acknowledgements have been received from the fda.Integra is resubmitting all impacted mdr reports for the time period 05nov2019 through 30jun2020.Integra lifesciences contacted (b)(4) office of product evaluation and quality on july 8-9, 2020 to report these issues regarding mdr reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key10277713
MDR Text Key203708113
Report Number3004608878-2020-00092
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 01/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Was Device Available for Evaluation? No
Date Manufacturer Received03/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MAYFIELD SKULL PINSREF (B)(4)
-
-