A physician reported that on (b)(6) 2020, the a3059 mayfield composite series skull clamp slipped during a cerebellum tumor surgery.The patient sustained a small laceration and hematoma that needed treatment.The physician reported that the injury could have been caused by having more pressure to the skull pins, and as the pressure was relieved, the skull clamp was not tightened down.Additional information has been requested.
|
Additional information: d10, g4, g7, h2, h3, h6, h10.Udi # (b)(4).The reported complaint was not confirmed as the product was not returned for evaluation.No dhr review was possible as no other lot or serial number was provided during the course of the complaint investigation.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|