• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED Back to Search Results
Model Number 71118200
Device Problem Material Fragmentation (1261)
Patient Problems No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/25/2020
Event Type  malfunction  
Event Description
It was reported that the tip of the flexible shaft w/cir connector is cracked and needs replacement.It is unknown whether the event happened during surgery and if there was patient involvement.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident this device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further investigation warranted for these complaints; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE SHAFT W/CIR CONNECTOR
Type of Device
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10280053
MDR Text Key199043646
Report Number1020279-2020-03236
Device Sequence Number1
Product Code HSD
UDI-Device Identifier03596010436528
UDI-Public03596010436528
Combination Product (y/n)N
PMA/PMN Number
K121714
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71118200
Device Catalogue Number71118200
Device Lot Number13JM10934
Was Device Available for Evaluation? No
Date Manufacturer Received04/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-