MAUDE Adverse Event Report: ST. JUDE MEDICAL PACEL BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PACEL BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number 401771
Device Problem Material Too Rigid or Stiff (1544)
Patient Problem Cardiac Perforation (2513)
Event Date 06/11/2020
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Two days post procedure, a right ventricle perforation was noted.During a transcatheter aortic valve implantation (tavi) procedure, the device was placed in the right ventricle.Two days following the procedure, the patient was transferred to the intensive care unit, where a right ventricle perforation was noted.The patient experienced rapid hypotension and a short cardiac arrest after the device was removed.An echocardiogram revealed a pericardial tamponade for which a pericardiocentesis was performed to stabilize the patient.The physician stated the pacing catheter felt hard following removal.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACEL BIPOLAR PACING CATHETER
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345 2126
MDR Report Key10282665
MDR Text Key199068929
Report Number2182269-2020-00062
Device Sequence Number1
Product Code LDF
UDI-Device Identifier05414734001137
UDI-Public05414734001137
Combination Product (y/n)N
PMA/PMN Number
K152784
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2023
Device Model Number401771
Device Catalogue Number401771
Device Lot Number7443703
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-