Sample was not received for evaluation.Dhr review did not identify any discrepancy during the product¿s manufacturing process which could have contributed and/or be related to the reported condition.The reported lot complied with all specifications and testing requirements.Failure analysis - retain samples from the same lot number were visually inspected; no anomaly was observed in any of the units; packaging was in good condition; sealing area was acceptable; no foreign matter was observed.Five (5) units were evaluated for thickness testing.Results: all tested units were within thickness specifications (0.01 to 0.03 in.).The complaint is considered unconfirmed, no further analysis is possible at the moment; therefore, the root cause is undetermined.
|