MAUDE Adverse Event Report: WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM Back to Search Results
Catalog Number 8065990739
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/26/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A health professional reported a secondary energy message during refractive treatment.The procedure was completed successfully.Additional information received; the patient did not experience any issues as a result of this event.There are two related reports for this event.This represents patient (b)(6) and an additional report will be filed.
 
Manufacturer Narrative
A review of the device history record traceable to the reported serial number indicates that the product was processed and released according to the product¿s acceptance criteria.Logfile review shows multiple error messages can be found in logfiles.The messages might occur if a gas change is needed and during the evacuation of the lines if the pressure does not fall below a set limit.A gas change was performed.During technical onsite visit the field service engineer (fse) replaced a sensor, beam splitter and performed system verification as per sir (service installation record).Root cause could not be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
MDR Report Key10286932
MDR Text Key199880172
Report Number3003288808-2020-00424
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065990739
Was Device Available for Evaluation? Yes
Date Manufacturer Received09/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age31 YR
-
-