HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER 2.0; VENTRICULAR (ASSIST) BYPASS
|
Back to Search Results |
|
Model Number 1420 |
Device Problems
Break (1069); Loose or Intermittent Connection (1371)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Investigation of this event is pending and a supplemental report will be sent upon its completion.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the controller had a loose component and a broken pin at the power port.The controller was exchanged.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for device evaluation.Product event summary: the controller was returned for evaluation.Failure analysis of the returned controller revealed damaged pins within power port one (1) of the controller.The damaged pins did not allow a battery to properly connect to the controller.No loose components or other anomalies were observed during failure analysis.As a result, the reported "broken pin at the power port¿" event was confirmed.The reported "loose component" event could not be confirmed; it is possible that the damaged pins were perceived as a loose component.The most likely root cause of the reported "broken pin" event can be attributed to a misalignment between the power port and battery output connector.Capa: pr00384004 was opened to investigate bent / damaged pins with controller 2.0.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for additional information.Additional information was received regarding the recall number.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|