• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC CLARION IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC CLARION IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number AB-5100H-01A
Device Problems Loose or Intermittent Connection (1371); Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2020
Event Type  malfunction  
Event Description
The recipient's device was reportedly explanted.Advanced bionics is in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.Additional information obtained reportedly indicates that the reason for explant was loss of lock.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Additional information: sections h.4, d.4, & d.10 disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.No further details will be provided.The external visual inspection revealed the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical test performed.The device passed the tests performed.This older device configuration is not currently manufactured.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The internal visual inspection revealed that the analog chip was damaged.This was induced during the failure analysis process.The reported complaint of loss of lock could not be confirmed due to damage induced to the hybrid during the failure analysis process.This device passed the limited electrical tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLARION IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10292399
MDR Text Key199446776
Report Number3006556115-2020-00630
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/17/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/01/2004
Device Model NumberAB-5100H-01A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age19 YR
-
-