• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER SITUATE; COUNTER, SPONGE, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER SITUATE; COUNTER, SPONGE, SURGICAL Back to Search Results
Model Number 01-0031
Device Problem False Positive Result (1227)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2020
Event Type  malfunction  
Manufacturer Narrative
This product is not sold in us and is 510(k) exempt.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a procedure, staff scanned patient with mat and the system detected a sponge.Count was correct.Staff rescanned with wand and no sponge was detected and they rescanned with mat once again and not sponge detected.There was no patient injury.
 
Manufacturer Narrative
Additional information: a1, a2, a3, a4(weight in lbs.), b5, d3(mfr name, street 1, mfr city, mfr region, postal code), g4.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during breast lumpectomy and sentinel node biopsy, staff scanned patient with mat and the system detected a sponge.Count was correct.Staff rescanned with wand and no sponge was detected and they rescanned with mat once again and no sponge detected.Nurse continued trying until she obtained a negative results on both matt and the wand.There was no patient injury.
 
Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device gave a false positive detection.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITUATE
Type of Device
COUNTER, SPONGE, SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder,co CA 80301 3299
MDR Report Key10294336
MDR Text Key201871126
Report Number3005883396-2020-00033
Device Sequence Number1
Product Code LWH
UDI-Device Identifier10884521700642
UDI-Public10884521700642
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number01-0031
Device Catalogue Number01-0031
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2021
Date Manufacturer Received04/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age76 YR
Patient Weight82
-
-