• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA AORTIC BIOPROSTHESIS; HEART-VALVE, REPLACEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA AORTIC BIOPROSTHESIS; HEART-VALVE, REPLACEMENT Back to Search Results
Model Number 3000
Device Problems Degraded (1153); Gradient Increase (1270); Appropriate Term/Code Not Available (3191)
Patient Problems Aortic Valve Stenosis (1717); Stenosis (2263)
Event Date 06/26/2020
Event Type  Injury  
Manufacturer Narrative
Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient related contributing factors.Structural valve deterioration (svd), a common reason for bioprosthesis explant or reoperation, encompasses multiple failure modes, including calcific and non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors, and is not usually an indication of a device malfunction.The device was not returned for evaluation, as it remains implanted.The root cause of this event cannot be conclusively determined.However, it is likely that patient related factors and the progression of the underlying valvular disease pathology contributed to the event.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
It was reported that a 23mm aortic valve was disabled via a valve-in-valve procedure after an implant duration of 7 years, 8 months due to stenosis.A 23mm transcatheter valve was implanted.As reported, the md did not allege any early failure of the surgical valve.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Manufacturer Narrative
H10: additional manufacturer narrative updated b5, b7, f10, and h6 per new information received.
 
Event Description
It was reported that a 23mm aortic valve was disabled via a valve-in-valve procedure after an implant duration of 7 years, 8 months due to degeneration and severe stenosis.A 23mm transcatheter valve was implanted.Patient was discharged on pod #1.As reported, the md did not allege any early failure of the surgical valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA AORTIC BIOPROSTHESIS
Type of Device
HEART-VALVE, REPLACEMENT
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
MDR Report Key10304662
MDR Text Key200027948
Report Number2015691-2020-12614
Device Sequence Number1
Product Code DYE
Combination Product (y/n)N
PMA/PMN Number
P860057/S018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/15/2015
Device Model Number3000
Was Device Available for Evaluation? No
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age67 YR
-
-