The device was tested in follow-up of the event and did not exhibit any malfunction; a self-test was passed without deviations and a ventilation episode with a test lung could be performed w/o problems.The evaluation of the electronic log file revealed that the concerned procedure was suffering from the beginning from a disturbed ventilation.The device alarmed repeatedly for apnea, airway pressure high and insp co2 high.After about 40 minutes the device started to alarm for a fresh gas deficit, in the following the ventilator was not able anymore to apply the set tidal volumes.Another 7 minutes later the device measured negative and positive pressure peaks (<-20mbar, +70mbar) while the ventilator piston was blocked at the upper end position.The workstation responded with a shutdown of automatic ventilation.Dräger finally concludes that there is no issue with the device which would require repair or correction.The disturbances of the ventilation were brought to the user's attention by corresponding alarms.The fresh gas deficit worsened over several minutes and led to a blocking of the ventilator piston at the end wherupon the device responded with shutdown of automatic ventilation.The origin of the massive pressure fluctuations that led to the blocking cannot be derived from the available information; imaginable would be spontaneous breathing of the patient or the use of a bronchial suction system.
|